Add to Cart
ST-115CS Fully automatic Kjeldahl nitrogen analyzer
Name: Automatic Kjeldahl nitrogen analyzer
model: ST-115CS=ST115C+STZ-08C
ST-115CS Kjeldahl nitrogen determination method is a classic method for determining nitrogen. The Kjeldahl method is used to determine soil, food, animal husbandry, agricultural products, feed, and other nitrogen-containing compounds. When using this method to determine samples, three processes are required: sample digestion, distillation separation, and titration analysis
The ST-115CS fully automatic Kjeldahl nitrogen analyzer is a "nitrogen element" (protein) analysis system developed based on the classic Kjeldahl nitrogen analyzer method, using fully automatic distillation titration technology and color method to determine the endpoint. Its method and manufacturing comply with the manufacturing standards of "GB/T 33862-2017 Fully (Semi) Automatic Kjeldahl Nitrogen Analyzer" and relevant international standards
Product Features:
●The operating system adopts a 7-inch color touch screen, with easy human-machine dialogue operation and easy learning
●The three-level permission management, electronic records, electronic tags, and operation traceability query system meet the relevant certification requirements
●The system automatically shuts down after 60 minutes of unmanned operation, which is energy-saving, safe, and reassuring
● Choose R, G, B coaxial three color light sources and three color sensors, with a wide range of color adaptability and high accuracy
●Automatic adjustment of R, G, and B light intensity suitable for analyzing samples of different concentrations
●The three titration curves of R, G, and B can observe the changes in the sample during the titration process, providing assistance for analysis
●The titration cup is installed in an open manner, making it convenient for users to observe the titration process and clean the titration cup
●Imported linear motor and injection needle tube form a "plunger titration pump" with high control accuracy and stability
●The titration speed can be set arbitrarily from 0.05ml/second to 1.0m/s
●Distillation time can be freely set from 10 seconds to 9990S
●The instrument has a built-in protein coefficient query table for users to access, query, and participate in system calculations, making it convenient for users to use
●16M data storage capacity can store 1 million sample data for reference, 5.7CM automatic paper cutting thermal printer
●99 groups of users set up their own program (recipe) management system
● Measurement of "nitro" class samples through delay analysis
●The metal material (304 stainless steel) steam pot has stable, reliable, and pressure resistant performance
● Leakage protection system prevents leakage and electric shock to ensure the safety of operators
●The digestive system does not add reagents or heat to prevent injury from reagents or steam
●The safety door is not closed and the system does not work. The 5mm thick transparent safety door ensures personal safety
●Automatic shutdown and alarm of steam system water shortage system to prevent accidents from occurring
●Automatic shutdown and alarm of steam boiler overheating system
TECHNICAL INDEX:
Analysis range | 0.1-240 mg N |
Precision (RSD) | ≤ 0.5% |
Recovery rate | 99-101% (± 1%) |
Titration accuracy | 0.2 μ L/step |
Titration speed | 0.05-1.0ml/s can be set arbitrarily |
Distillation time | 10-9990S, freely adjustable |
Sample analysis time | 4-6 minutes/(cooling water temperature 18 ℃) |
Concentration range of titrant | 0.01-5 mol/L |
Titration cup volume | 300ml |
Data storage capacity | 1 million sets of data |
Printer | 5.7CM thermal automatic paper cutting printer |
Delay analysis | reaction time 0-2 hours |
Delay alkali addition time | 0-999 seconds |
Automatic shutdown time | 60 minutes |
Working voltage | AC220V/50HZ |
Heating power | 2000W |
Dimensions | 500 * 460 * 710mm |
Net weight | 36Kg |
●Analysis range:0.1-240 mg N
●Precision (RSD):≤ 0.5%
●Recovery rate:99-101% (± 1%)
●Titration accuracy:0.2 μ L/step
●Titration speed:0.05-1.0ml/s can be set arbitrarily
●Distillation time:10-9990S, freely adjustable
●Sample analysis time:4-6 minutes/(cooling water temperature 18 ℃)
●Concentration range of titrant:0.01-5 mol/L
● Titration concentration input method: manual input
●Titration cup volume:300ml
●Data storage capacity:1 million sets of data
●Printer:5.7CM thermal automatic paper cutting printer
●Communication interface:refrigeration cycle unit
●Delay analysis:reaction time 0-2 hours
●Delay alkali addition time:0-999 seconds
●Automatic shutdown time:60 minutes
●Working voltage:AC220V/50HZ
●Heating power:2000W
● Dimensions:500 * 460 * 710mm
●Net weight:36Kg
●Product Configuration Checklist
●3 reagent tanks, 1 waste liquid tank, and 4 reagent pipelines
●2 cooling water pipes, 1 power cord, and 1 M19 open handle
●8 rolls of 5.7cm printing paper and 1 1L titration reagent bottle
●STZ-08C digital precision digestion furnace:
Product features:
1.The heating element (module) is made of aluminum alloy precision casting and processing, which is acid and alkali resistant and has a long service life
2.The furnace hole temperature is continuously adjustable, with a fast heating rate and a wide range of applications
3.The temperature difference between the pores is small, the sample digestion consistency is good, the thermal efficiency is high, and it is conducive to the digestion of the sample
4.The instrument table and test tube rack are welded with brushed stainless steel plates, which have a certain anti-corrosion ability
5.Adopting a new generation LCD temperature controller with functions such as timed shutdown and alarm, PID intelligent control technology, high temperature control accuracy, small temperature fluctuation amplitude, and easy to learn
6.Optional 30 segment program heating function to meet users' various experimental needs
Parameter indicators:
Number of heating hole | 8 |
Volume | 250ml |
Design temperature | RT--500℃ |
Temperature control method | LCD display, intelligent PID control |
Temperature control accuracy | ± 1 ℃ |
Heating rate | ≈ 20℃/min |
Voltage | AC220V |
Power | 2KW |
Current | 9A |
Heating elemen | Nickel chromium wire aluminum alloy heating plate |
1.Number of heating holes:8
2.Volume:250ml
3.Design temperature:RT --500 ℃
4.Temperature control method:LCD display, intelligent PID control
5.Temperature control accuracy:± 1 ℃
6.Heating rate:≈ 20℃/min
6.Voltage:AC220V
7.Power:2KW
8.Current:9A
9.Heating element:Nickel chromium wire aluminum alloy heating plate